全髋关节置换术后股骨偏心距与聚乙烯磨损的相关性研究Correlation between femoral offset and polyethylene wear after total hip arthroplasty
聂志刚;彭昊;陈森;鲁强;邓爽;周观金;
摘要(Abstract):
[目的]探讨人工全髋关节置换术后股骨偏心距与聚乙烯磨损的相关性。[方法]用计算机辅助测量方法测量髋关节置换聚乙烯内衬的线性磨损。以股骨头中心参照于髋臼杯开口面轴线的位移,计算线性磨损的大小与方向。对2008年11月~2015年10月本院骨科78例使用非骨水泥型金属对聚乙烯假体患者的X线片进行股骨偏心距和聚乙烯磨损的测量,平均随访时间4.08年。[结果]平均线性磨损率(0.18±0.04)mm/年,平均线性磨损方向与髋臼杯轴线的夹角(33.0±18.2)°。术中股骨偏心距得到重建组的线性磨损率小于未得到重建组(P<0.001)。股骨偏心距得到重建组与未得到重建组的磨损方向差异无统计学意义(P=0.210)。对股骨双侧偏心距差值与年磨损率进行相关分析,两者成正相关性(r=0.669,P<0.001)。[结论]人工全髋关节置换术中重建股骨偏心距可减慢聚乙烯内衬磨损速度,延长髋关节假体的寿命,改善远期效果。
关键词(KeyWords): 关节成形术;置换;髋;聚乙烯;偏心距
基金项目(Foundation):
作者(Author): 聂志刚;彭昊;陈森;鲁强;邓爽;周观金;
Email:
DOI:
参考文献(References):
- [1]兰天,肖军,史占军.髋关节假体聚乙烯磨损及其影响因素的相关性分析[J].中华医学杂志,2014,94(43):3416-3420.
- [2]Mahmood SS,Mukka SS,Crnalic S,et al.Association between changes in global femoral offset after total hip arthroplasty and function,quality of life,and abductor muscle strength.A prospective cohort study of 222 patients[J].Acta Orthop,2016,87(1):36-41.
- [3]Davey JR,O’Connor DV,Burke DW。Femoral component offset.Its effect on strain in bone-cement[J].J Arthroplasty,1993,8(1):23-26.
- [4]Charles MN,Bourne RB,Davey JR,et al.Soft-tissue balancing of the hip:the role of femoral offset restoration[J].Instr Course Lect,2005,54(2):131–141.
- [5]Little NJ,Busch CA,Gallagher JA,et al.Acetabular polyethylene wear and acetabular inclination and femoral offset[J].Clin Orthop,2009,467(11):2895-2900.
- [6]肖瑜,张福江,马信龙,等.全髋关节置换术后股骨偏距减小与聚乙烯内衬磨损的关系[J].中华骨科杂志,2012,32(9):849-854.
- [7]Dorr LD,Wan Z.Ten years of experience with porous acetabular components for revision surgery[J].Clin Orthop,1995,319(319):191-200.
- [8]Christodoulou NA,Dialetis KP,Christodoulou AN.High hip center technique using a biconical threaded Zweymuller cup in osteoarthritis secondary to congenital hip disease[J].Clin Orthop,2010,468(7):1912-1919.
- [9]芦北极,孙俊英,施勤,等.全髋关节成形术后的偏心距测量对临床功能的影响[J].中国矫形外科杂志,2012,20(14):1282-1286.
- [10]Lechler P,Frink M,Gulati A,et al.The influence of hip rotation on femoral offset in plain radiographs[J].Acta Orthop,2014,85(4):389-395.
- [11]Callaghan JJ,Pedersen DR,Johnston RC,et al.Clinical biomechanics of wear in total hip arthroplasty[J].Iowa Orthop J,2003,23(1):1-12.
- [12]Mc Calden RW,Naudie DD,Yuan X,et al.Radiographic methods for the assessment of polyethylene wear after total hip arthroplasty[J].J Bone Joint Surg Am,2005,87(10):2323–2334.
- [13]Callaghan JJ,Pedersen DR,Olejniczak JP,et al.Radiographic measurement of wear in 5 cohorts of patients observed for 5 to 22 years[J].Clin Orthop,1995,(317):14-18.