单双球囊后凸成形术对老年性椎体骨折的实验研究Ex vivo comparative study between single and double balloon bilateral dilatation of kyphoplasty
郝润松;孙钢;于志军;刘训伟;金鹏;李凡东;谢志勇;易玉海;
摘要(Abstract):
[目的]探讨单、双球囊后凸成形术对治疗老年骨质疏松性椎体压缩骨折的差别。[方法]10具老年胸腰段脊柱标本(T12~L2),骨密度仪测定椎体骨密度,证实为骨质疏松标本。标本随机分配为单球囊组和双球囊组,每组5具。将标本游离成单个椎体,每具标本选择2个椎体共20个椎体,经万能力学试验机轴向加载,制造椎体压缩骨折,压缩程度为25%,随即行单、双球囊后凸成形术,测定椎体在骨折前、后及骨水泥注入后的高度并记录骨水泥注入前、后最大负荷,计算其刚度。[结果]单球囊组原始椎体最大压缩负荷及刚度分别为(1420±80.55)N和(1030.27±50.74)N/mm,骨水泥注入后分别为(2937.9±175.15)N和(963±20.66)N/mm,双球囊组原始椎体最大负荷及刚度分别为(1398±85.35)N和(1024.16±57.50)N/mm,骨水泥注入后分别为(2800.8±157.79)N和(964.52±22.03)N/mm。单球囊组骨水泥注入后椎体高度恢复率为(96.20±0.01)%;双球囊组骨水泥注入后椎体高度恢复率为(95.30±0.02)%。骨水泥注入后椎体最大负荷及对骨折椎体的高度恢复单球囊组与双球囊组间差异无显著性(P<0.01),椎体刚度差异无统计学意义。两组均未见骨水泥溢出。[结论]单、双球囊椎体后凸成形术生物力学测量结果和椎体高度恢复等实验结果无显著性统计学差异且骨水泥的渗漏率极低。
关键词(KeyWords): 骨质疏松;椎体压缩性骨折;球囊;椎体后凸成形术
基金项目(Foundation):
作者(Author): 郝润松;孙钢;于志军;刘训伟;金鹏;李凡东;谢志勇;易玉海;
Email:
DOI:
参考文献(References):
- [1]Eriksson SA,Isberg BO,Lindgren JU.Prediction of vertebral strength by dual phonton absorptiometry and quantitative computed tomography[J].Calciftissue Int,1989,44:243-250.
- [2]Tae HL,Gregory T,Brebach,et al.Biomechanical evaluation of an injectable calcium phosphate cement for vertebroplasty[J]Spine,2002,12:1297-1302.
- [3]Sun G,Cong YJ,Xie ZG,et al.Percutaneous vertebroplasty using instruments and drugs made in China for vertebral metastasis[J].Chin Med J,2003,116:1207-1212.
- [4]孙钢,金鹏,易玉海,等.用国产器械行椎体后凸成形术治疗老年骨质疏松脊柱压缩骨折的价值[J].中华放射学杂志,2006,40:1089-1092.
- [5]Kado DM,Browner WS,Palermo L,et al.Vertebral fracture and mortality in older women:a prospective study[J].Arch Intern Med,1999,159:487-492.
- [6]Lieberman IH,Dudeney S,Reinhardt MK,et al.Initial outcome and eficacy of“kyphoplasty”in the treatment of painful osteoporotic vertebral compression fractures[J].Spine,2001,26:1631-1638.
- [7]Galibert P,Deramond H,Rosat P,et al.Preliminary note on the treatment of vertebral angioma by percutaneous acylic vertebroplasty[J].Neurochirurgie(French),1987,33:166-168.
- [8]孙钢,金鹏,刘训伟,等.单球囊体后凸成形术治疗骨质疏松性脊柱压缩骨折[J].医学影像学杂志,2007,17:325-328.
- [9]Belkoff SM,Mathis JM,Fenton DC,et al.An ex vivo biomechanical evaluation of an inflatable bone tamp used in the treatment of com-pression fracture[J].Spine,2001,2:151-156.